DS-71:
DS-72:
Features
Click here for the dedicated LAQUA website (opens in a new Window)
LAQUA DS-70 series measures Conductivity, Resistivity, Salinity, TDS and Temperature simultaneously and the Model DS-72 complies with industrial pharmacopoeias standard (JIS, USP, EP, JP and CP).
LAQUA DS-70 series measures Conductivity, Resistivity, Salinity, TDS and Temperature simultaneously and the Model DS-72 complies with industrial pharmacopoeias standard (JIS, USP, EP, JP and CP).
The meters’ auto diagnostic function allows the user
to make sure that the instruments will always give high performance and
reliable measurements. Fully GLP/GMP compliant, the LAQUA can be
connected to an external printer. To ensure traceability all results are
recorded together with sample name, user ID and calibration
information.
We offer a comprehensive range of conductivity
electrodes adapted to all type of application and sample. Recognised for
their stability and longevity, HORIBA’s electrodes provide accurate
measurement in all conditions.
LAQUA’s large capacitance touch screen panel:
- Display easy to understand value and functions menus
- Switch between digital real time graphs or analogue displays during measurement with just the flick of a finger.
- Full on-screen support on the meter via an electronic user guide with colour pictures that can be called up at any time, featuring problem solver guide for calibration or measurement.
Functions
- Easy navigation for main unit and electrode inspections.(DS-72)
- Full support for various country pharmaceutical pure water guidelines(USP/EP/JP/CP)(DS-72)
- Customizable auto-hold function for calibration and measurement (DS-72)
- Simultaneous connection to a GLP/GMP compatible and PC
- Degital memory:Maximum 2,000 sets of measurement data can be recorded (DS-71:999)
- USB-PC communication (DS-71/72) and USB memory (DS-72)
- Multi-language support(Japanese, English, Chinese, Korean) (DS-72)
- FDA 21 CFR Part11 (Please ask for quotation)
Specifications
DS-71DS-72ConductivityMeasurement method2 AC bipolar methodMeasurement range
(Display range)Cell constant 100 m-1:0.000 mS/m~19.99 S/m
Cell constant 10 m-1:0.0 μS/m~1.999 S/m
Cell constant 1000m-1:0.00 mS/m~199.9 S/mResolution0.05% of full scaleRepeatability±0.5% ±1 digitChange unit○○Distilled water temperature conversion○○Periodical check×○P/EP/USP/CP Pharmaceutical water aplication×○SalinityMeasurement methodConversion from conductivity valueMeasurement range
(Display range)0.00~80.00 PPT(0.000%~8.000%)Resolution0.01 PPT(0.001%)Salt concentration calibration○○ResistivityMeasurement methodConversion from conductivity valueMeasurement range
(Display range)Cell constant 100 m-1:0.00 Ω・m~199.9 kΩ・m
Cell constant 10 m-1:0.0 Ω・m~1.999 MΩ・m
Cell constant 1000 m-1:0.000 Ω・m~19.99 kΩ・mResolution0.05% F.S.Repeatability±0.5%F.S.±1 digitTDSMeasurement methodConversion from conductivity valueMeasurement range
(Display range)0.01 mg/L ~100 g/L0.01 mg/L ~1000 g/LResolution0.01 mg/LTemperatureMeasurement range0.0~100.0℃(-30.0~130.0℃)Resolution0.1℃Repeatability±0.1℃±1 digitInput/outputInput(CH)11USB peripherals (Communication with PC)*1○○USB host(USB memory)×○RS-232C(Printer/PC)○○Analog out put×○DataMemory number9992,000Interval memory○○ID input○○Data search×○DisplayDisplayCustom LCDColor graphic LCD with capacitive Touch PanelMultilanguage display×Japanese/English
/Chinese/KoreanFunctionNavigation function×○User gide×○Graf display×○Printer connectivity(GLP/GMP)○○Custom printing function×○Temperature compensation(Auto/manual)○○AutoHold function○○AutoHold setting×○Register operator×○Security(pass word)○○Version up function○○Ambient temperature0 ℃ ~ 45℃PowerAC adaptor 100 ~ 240 V 50/60 HzDemensions170(W)×174(D)×73(H)mm(Excluding electrode stand and AC adaptor)Power consumptionApprox. 0.7 VAApprox. 9.8 VAMass of main unitApprox .500 gApprox. 700 g*1 USB cable sold separately. Software can be download by web registration.