The BINDER ED series

Selasa, 17 Februari 2015

 

 

 

 

 

 

Heating oven with natural convection

The strength of the BINDER ED series heating oven: routine drying and sterilization applications up to 300 °C. Thanks to the gravity convection, thermal processes are highly efficient for this heating oven. A BINDER ED heating oven ensures fast, uniform drying.

Advantages:

  • Fast, even tempering
  • Wide temperature range
  • Quality "Made in Germany"
Features Customer benefits Characteristics
APT.line™
  • Quick drying
  • Identical test conditions throughout the chamber interior
  • Independent of specimen size and quantity
APT.line™ for maximum precision
  • Uniform circulation even under full load
  • Homogeneous temperature conditions throughout specimen material
Temperature range
  • Broad range of applications
  • Short warm up times
Standard up to 300 °C
  • Large power reserves
Inner chamber concept
  • Maximum occupational safety
  • Easy loading and unloading of specimen material
  • Easy cleaning
Inner chamber made of stainless steel
  • Very tight door closure with 2-point door latch
  • Low heat dissipation due to 60 mm insulation
  • Rack with tilt protection
  • No permanent fixtures
Standard equipment
  • Reliable, easy handling
Comprehensive standards
  • Microprocessor control
  • Ergonomically positioned controller.
  • RS 422 interface
  • PT 100 temperature sensor
Quality
  • Reliable devices with long service lives
  • Short delivery times
  • Minimal maintenance and operating costs
Premium quality
  • Highly automated
  • Series production
  • High-quality materials, state-of-the-art production technology
  • High standard according to DIN 12880 (27-point measurement)
Accessories and Services
  • Flexible solution in terms of size, type and equipment
  • Optimal solution for numerous applications
  • BINDER INDIVIDUAL for customer-specific solutions
  • Worldwide BINDER Service
Comprehensive product portfolio
  • Size 23 to 720 liters
  • Additional product lines with humidity, light, CO2 or vacuum
  • Voltage variants (UL) and certificates
  • Various options: Door with viewing window, access ports, reinforced shelves, Data Logger Kits
  • Worldwide service network
    • Electronically controlled APT.line™ preheating chamber assuring temperature accuracy and reproducible results
    • Temperature range from 5 °C above ambient temperature to 300 °C
    • Digital temperature setting with an accuracy of one degree
    • DS control with integrated timer 0 to 99,59 hrs
    • Independent adjustable temperature safety device class 2 (DIN 12880), with visual temperature alarm
    • Adjustable ventilation by means of front ventilation flap slide and rear exhaust Ø 50 mm
    • Optional RS 422 Interface for APT-COM™ Data Control System communication software
    • Units up to 115 liters are stackable
    • 2 chrome-plated racks included
    • ED 23 with space-saving door mechanism
    • BINDER test confirmation

      ED 23
      Exterior dimensions

      Housing

      Width (mm) 435

      Height (incl. feet/castors) (mm) 495

      Depth (plus connections and fittings) (mm) 520

      Depth (incl. connections and fittings) (mm) 620

      Wall clearance, rear (mm) 100

      Wall clearance, side (mm) 100

      Exhaust duct (outer Ø mm) 52

      Steam space volume (l) 36

      Number of doors (ea.) 1
           
      Interior dimensions

      Width (mm) 222

      Height (mm) 330

      Depth (mm) 300

      Interior volume (l) 20

      Racks (number standard/max.) 2 / 3

      Load per rack (kg) 12

      Permitted total load (kg) 25

      Weight (empty) (kg) 22
           
      Temperature data

      Temperature range approx. 5 °C above ambient temperature to (°C) 300

      Temperature variation 1)

      at 70 °C (± K) 1,5

      at 150 °C (± K) 2,5

      at 300 °C (± K) 3,8

      Temperature fluctuation at 70 °C (± K) 0,3

      Warm-up time 2)

      to 70 °C (min.) 13

      to 150 °C (min.) 24

      to 250 °C (min.) 35

      Recovery time after doors were open for 30 sec. 2)

      at 70 °C (min.) 2,5

      at 150 °C (min.) 5

      at 300 °C (min.) 8
           
      Ventilation data

      Ventilation

      at 70 °C (x/h) 10

      at 150 °C (x/h) 13

      at 300 °C (x/h) 17
           
      Electrical data

      IP protection class acc. to EN 60529 IP 20

      Voltage (± 10%) 50 / 60 Hz (V) 230

      Nominal power (kW) 0,8

      Energy consumption

      at 70 °C (W) 43

      at 150 °C (W) 148

      at 300 °C (W) 450
      1) without a viewing window
      2) To 98% of the set value

      All technical data are specified for units with standard equipment at an ambient temperature of +25 °C and a line voltage fluctuation of ±10%. The temperature data is determined in accordance to factory standard following DIN 12880, respecting the recommended wall clearances of 10% of the height, width and depth of the inner chamber. All figures are typical average values for series devices. We reserve the right to alter technical specifications at any time.
      Numerous access ports

      Numerous access ports

      With silicone plugs for introducing external measuring instruments into the chamber, access ports with 10, 30, 50 mm diameters.
      Door with window and interior lighting

      Door with window and interior lighting

      For optimal process control in the inner chamber, available for all device sizes.
      Door lock

      Door lock

      Prevents unauthorized access to the process sequences in the chamber.
      Calibration certificate & validation

      Calibration certificate & validation

      BINDER can significantly reduce the workload in qualifying and validating devices. Nobody knows our devices as well and has as much experience in certifications as we do.
      ED 23
      Access ports with silicone plug, 10, 30, 50, 100 mm O
      Anti-slip rubber pads for safe stacking (1 set of 4 pieces) O
      Switchable audible alarm for over temperature. Adjustable limit on the independent temperature safety device O
      Independent temperature safety device class 3.1 (DIN 12880) with optical alarm O
      Analog output for temperature 4 - 20 mA with 6-pin DIN socket (output not adjustable) O
      Temperature measurement acc. to DIN 12880 (27 measuring points, for 23 liter devices 15 measuring points) at 150 °C or at specified temperature with measuring protocol and certificate O
      Calibration certificate, measurement in center of chamber at 150 °C or at specified testing temperature O
      Extension to calibration certificate. Each additional measurement at additional measuring point or testing temperature O
      Data Logger Kit T 350: For continuous temperature recording of 0 °C to 350 °C. Kit includes 1 data logger, PT 100 sensor with 2 m Teflon extension cable and 1 fixture for mounting to the BINDER unit O
      Data Logger Software: Configuration and evaluation software for all BINDER Data Logger Kits, incl. data cable O
      Rack, chrome-plated O
      Rack, stainless steel O
      Perforated shelf, stainless steel O
      Door lock O
      Door gasket, FKM (Viton) O
      Door with window 180 x 180 mm and interior lighting (15 W) O

0 komentar:

Posting Komentar

Free Website templateswww.seodesign.usFree Flash TemplatesRiad In FezFree joomla templatesAgence Web MarocMusic Videos OnlineFree Wordpress Themeswww.freethemes4all.comFree Blog TemplatesLast NewsFree CMS TemplatesFree CSS TemplatesSoccer Videos OnlineFree Wordpress ThemesFree CSS Templates Dreamweaver